Low sensitivity with Xevo for clinical samples - WKB115909

Article number: 115909

SYMPTOMS

  • Low sensitivity

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-S micro

CAUSE

Argon flow set too low or empty argon cylinder

FIX or WORKAROUND

Adjust Argon pressure to recommended pressure (typically 7 psi).

OR

Replace the argon cylinder if needed

ADDITIONAL INFORMATION