Low sensitivity in both positive and negative modes on MS systems - WKB112525
SYMPTOMS
- Low sensitivity in both positive and negative modes
ENVIRONMENT
- ACQUITY QDa Mass Detector
- 3100 Mass Detector
- SQ Detector
- TQ Detector
- SQ Detector 2
- Xevo TQD MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT G2-XS
- Xevo G2-S Tof
- Xevo G2-XS Tof
CAUSE
Dirty or blocked sample cone or aperture disc
FIX or WORKAROUND
Clean or replace the sample cone/gas nozzle, or replace the aperture disk for the QDa.
Clean or replace Probe
ADDITIONAL INFORMATION
Ordering information:
QDa: Aperture Disc Assembly, 0.2mm, Performance (# 700009768), Sample Cone (# 700009597)
3100MS / SQD / TQD: Gas Cone, Camlock (# 700003054), Sample Cone, 0.5 Diam. (# 700003055)
SQD 2 / Xevo TQD: Cone Gas Nozzle (# 700004216), Sample Cone (# 700005997)
Xevo TQ MS: Cone Gas Nozzle (# 700004216), Sample Cone (# 700004219)
Xevo TQ-S: Cone Gas Nozzle (# 700004216), Sample Cone (# 700005350)
Xevo TQ-XS: Cone Gas Nozzle (# 700004216), Sample Cone (# 700011217)
Xevo TQ-S cronos: Cone Gas Nozzle (# 700004216), Sample Cone (# 700012275)
Xevo TQ-S micro: Cone Gas Nozzle (# 700004216),Sample Cone (# 700010289)
SYNAPT G2-S / Si: Cone Gas Nozzle (# 700004216), Sample Cone (# 700005350)
SYNAPT XS: Cone Gas Nozzle (# 700004216), Sample Cone (# 700011217)
Xevo G2-S / XS: Cone Gas Nozzle (# 700004216), Sample Cone (# 700009087)
id112525, A-SQ, A-SQ, MALDIG2SI, Q3100MSD, QDA, SQD, SQD2, SYNAPTXS, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD