"Instrument Voltage Readback Failure" error on ACQUITY QDa Mass Detector - WKB112302
Article number: 112302
SYMPTOMS
- "Instrument Voltage Readback Failure" error
- Fails after source pressure test
ENVIRONMENT
- ACQUITY QDa Mass Detector
- RADIAN ASAP Direct Mass Detector
CAUSE
The critical readback report indicates that the detector fails startup due to turbo power out of range (36.6 W and not 37.0 W).
FIX or WORKAROUND
Replace the aperture disk.
ADDITIONAL INFORMATION
For part numbers, see the article What are the part numbers for the source aperture carrier, aperture disk, and sample cone for an ACQUITY QDa? - WKB52732
For a video to demonstrate how to replace the aperture disk, click here