TQ Detector Cone Voltage readback is 0 V - WKB112188

Article number: 112188

SYMPTOMS

  • Cone Voltage Readback 0 V
  • Adjusting voltage has no impact
  • Changing ion mode has no impact

ENVIRONMENT

  • ACQUITY TQD System
  • TQ Detector

CAUSE

Unknown

FIX or WORKAROUND

  1. Put the TQ Detector in Standby.
  2. Reboot the EPC by pressing the red reset button, and wait for the TQD to re-initialize.
  3. Open the tune page, turn on the API gas, and put the TQD back into Operate.
  4. Check the cone voltage and observe functionality.

ADDITIONAL INFORMATION

 

id112188, TQD, TQDIVD