TQ Detector Cone Voltage readback is 0 V - WKB112188
Article number: 112188
SYMPTOMS
- Cone Voltage Readback 0 V
- Adjusting voltage has no impact
- Changing ion mode has no impact
ENVIRONMENT
- ACQUITY TQD System
- TQ Detector
CAUSE
Unknown
FIX or WORKAROUND
- Put the TQ Detector in Standby.
- Reboot the EPC by pressing the red reset button, and wait for the TQD to re-initialize.
- Open the tune page, turn on the API gas, and put the TQD back into Operate.
- Check the cone voltage and observe functionality.
ADDITIONAL INFORMATION
id112188, TQD, TQDIVD