Low desolvation and cone gas flow on the Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB108070
Article number: 108070
SYMPTOMS
- Low desolvation gas flow; readback is 29 L/hr with a set point of 550 L/hr
- Low cone gas flow; readback is 39 L/hr with a set point of 150 L/hr
- If API gas is turned off, desolvation and cone gas readbacks both go to 0 L/hr
- Restarting MassLynx does not resolve
- Rebooting the EPC does not resolve
- Rebooting instrument electronics does not resolve
ENVIRONMENT
- Xevo TQ-XS
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
CAUSE
Unknown
FIX or WORKAROUND
- Open and close the source door.
- Allow the source pressure test to run.
ADDITIONAL INFORMATION
id108070, XEVOG2XSQT, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD