Low desolvation and cone gas flow on the Xevo TQ-S, Xevo TQ-S cronos, Xevo TQ-S micro or Xevo TQ-XS - WKB108070

Article number: 108070

SYMPTOMS

  • Low desolvation gas flow; readback is 29 L/hr with a set point of 550 L/hr
  • Low cone gas flow; readback is 39 L/hr with a set point of 150 L/hr
  • If API gas is turned off, desolvation and cone gas readbacks both go to 0 L/hr
  • Restarting MassLynx does not resolve
  • Rebooting the EPC does not resolve
  • Rebooting instrument electronics does not resolve

ENVIRONMENT

  • Xevo TQ-XS
  • Xevo TQ-S
  • Xevo TQ-S cronos
  • Xevo TQ-S micro

CAUSE

Unknown

FIX or WORKAROUND

  1. Open and close the source door.
  2. Allow the source pressure test to run.

ADDITIONAL INFORMATION

 