Xevo G2-XS Tof vents after maintenance on the sample cone and cone gas nozzle - WKB106911
Article number: 106911
SYMPTOMS
- Replaced cone and cone gas nozzle after performing maintenance
- Turbo pumps start to slow down after opening isolation valve
- Instrument vents when turbo speed slows too much
- Cannot restart vacuum system
- Vacuum menu options show both Pump and Vent
ENVIRONMENT
- Xevo G2-XS Tof
CAUSE
Cone and cone gas nozzle do not seat properly when the isolation valve is opened.
FIX or WORKAROUND
- On the Tune page, go to the Vacuum menu and select VENT.
- Click YES when asked if you want to vent the mass spectrometer.
- Remove the cone and cone gas nozzle, leaving the isolation valve open.
- Reseat the cone and cone gas nozzle, leaving the isolation valve open.
- Go to the Vacuum menu and select pump.
- Allow the vacuum system to pump down.
ADDITIONAL INFORMATION
This issue and the resolution applies to all mass spectrometers except the ACQUITY QDa Mass Detector and the Xevo TQ-GC.