Poor chromatography/low sensitivity on the Xevo TQ-XS using APGC - WKB106788

Article number: 106788

SYMPTOMS

Reduced signal on SST injections.   

Chromatography poor.

ENVIRONMENT

  • Xevo TQ-XS
  • APGC

CAUSE

Contaminated corona pin or cone.

FIX or WORKAROUND

  1. Clean or replace corona pin (Corona Discharge Pin Assembly (700004809).
  2. Optimise corona pin position.
  3. Clean cone (TQXS Cone 700011217).

ADDITIONAL INFORMATION

 

id106788, APGC, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD