Poor chromatography/low sensitivity on the Xevo TQ-XS using APGC - WKB106788
Article number: 106788
SYMPTOMS
Reduced signal on SST injections.
Chromatography poor.
ENVIRONMENT
- Xevo TQ-XS
- APGC
CAUSE
Contaminated corona pin or cone.
FIX or WORKAROUND
- Clean or replace corona pin (Corona Discharge Pin Assembly (700004809).
- Optimise corona pin position.
- Clean cone (TQXS Cone 700011217).
ADDITIONAL INFORMATION
id106788, APGC, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD