Article number: 106788

SYMPTOMS

Reduced signal on SST injections.   

Chromatography poor.

Drops in intesity between injection caused by 'charging' on cone

ENVIRONMENT

  • Xevo TQ-XS
  • APGC

CAUSE

Contaminated cone.

FIX or WORKAROUND

  1. Clean cone (TQXS Cone 700011217)

ADDITIONAL INFORMATION

 

id106788, APGC, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD