Incorrect cone voltage readback on Xevo TQ-S and Xevo TQ-S micro - WKB10652

Article number: 10652

SYMPTOMS

  • Cone voltage readback = 94 V  (set to 14 V)

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-S micro

CAUSE

This is normal.

FIX or WORKAROUND

  1. The cone voltage readback is the sum of the voltages on several components: Source offset, Stepwave body gradient, Stepwave head gradient, and Collision energy, in addition to the cone voltage setting.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.