ESI capillary fell out - WKB1065
Article number: 1065
SYMPTOMS
- The capillary is visibly protruding from the ESI probe by more than 1 cm.
ENVIRONMENT
- Xevo TQ
- Xevo TQ-S
- Xevo TQD
- Xevo TQ-S micro
- Xevo TQ XS
- Xevo QTof
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Synapt G2
- Synapt G2-S
- Synapt G2-Si
- SQ Detector 2
CAUSE
The capillary is dislodged from the union in the ESI probe.
FIX or WORKAROUND
Rebuild the probe by following the steps on page 175 of the Xevo G2-XS QTof Overview and Maintenance Guide
The procedure is the same for all instruments mentioned in the environment section above. For quadrupole instruments, ignore any reference to the LockSpray.
ADDITIONAL INFORMATION
The probe may need to be rebuilt for other reasons, such as blockages or a bent probe tip. The same procedure applies.
id1065, A-SQ, Electrospray, MALDIG2SI, SQD2, SYNG2SIHD, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVTQSMIVD