SYMPTOMS

  • Removed the line from the valve, but the ferrule is still stuck inside the valve
  • Fishing out the ferrule with PEEK tubing doesn't work

ENVIRONMENT

  • Mass spectrometer
  • ACQUITY

CAUSE

The ferrule is jammed into the valve port.

FIX or WORKAROUND

  1. Use a Phillips-head screwdriver about the size of the PEEK tubing.
  2. Push the screwdriver into the hole in the PEEK ferrule.
  3. Once lodged, turn and pull the ferrule out. 

ADDITIONAL INFORMATION

id10632,

Not able to find a solution? Click here to request help.