Xevo TQ-S or Xevo TQ-XS: "Leak Detected" error - WKB100795

Article number: 100795

SYMPTOMS

  • "Leak Detected" error message
  • The trough is dry underneath the diverter and switching valves
  • The trough under the fluidics lines is full of solvent

ENVIRONMENT

  • Xevo TQ-S
  • Xevo TQ-XS
  • All Xevo Mass Spectrometers

CAUSE

The solvent waste line is back-filling from the waste bottle.

FIX or WORKAROUND

Cut the waste tubing, if necessary, and correctly route the waste line from the Mass Spectrometer to ensure that solvent drains properly. 

ADDITIONAL INFORMATION

Waste container connection.PNG

Figure 1:waste container connection.

id100795, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.