Xevo TQ-S or Xevo TQ-XS: "Leak Detected" error - WKB100795
Article number: 100795
SYMPTOMS
- "Leak Detected" error message
- The trough is dry underneath the diverter and switching valves
- The trough under the fluidics lines is full of solvent
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
- All Xevo Mass Spectrometers
CAUSE
The solvent waste line is back-filling from the waste bottle.
FIX or WORKAROUND
Cut the waste tubing, if necessary, and correctly route the waste line from the Mass Spectrometer to ensure that solvent drains properly.
ADDITIONAL INFORMATION
Figure 1:waste container connection.
id100795, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD