Xevo TQ-XS shows a high baseline - WKB100101
Article number: 100101
SYMPTOMS
- The baseline is about 100 times higher than when it was normal
- The baseline is high even if the liquid flow is stopped
ENVIRONMENT
- Xevo TQ-XS
- ACQUITY UPLC H-Class Bio
CAUSE
Dirty source enclosure
FIX or WORKAROUND
Wipe the source enclosure and the probe adjuster with methanol.
ADDITIONAL INFORMATION
id100101, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD