Sample manager leak detected - WKB99611
Article number: 99611
SYMPTOMS
- Error message: Sample manager leak detected
- Every two hours the leak sensor on the sample manager gives error message
ENVIRONMENT
- ACQUITY UPLC I-Class
- ACQUITY UPLC Column Manager
CAUSE
A loose fitting in the column manager inlet or outlet causes leaks that trigger the leak sensor.
FIX or WORKAROUND
- Tighten the fitting for the capillary between the sample manager and the column manager.
- Dry out the leak sensor and the cup that it sits in and reset the module.
ADDITIONAL INFORMATION
id99611, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+