Skip to main content
Waters

High area RSD for ACQUITY UPLC Sample Manager - WKB99615

Article number: 99615

SYMPTOMS

  • High area %RSD
  • New injection valve and sample loop
  • The volume detection device capillary is not crushed
  • Sample syringe leak test passes
  • No leaks

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

CAUSE

Bent puncture needle

FIX or WORKAROUND

Replace the puncture needle (700003898).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.