On 6890 Plus with manual injection, cannot acquire data without error - WKB99522
Article number: 99522
SYMPTOMS
- 6890 Plus manual injection cannot acquire data without error
- ALS injection acquires data normally
ENVIRONMENT
- Agilent 6890 GC
- Empower 3 Feature Release 4 Service Release 3 Hotfix 1
- Agilent ICF Support v3.1
- Agilent ICF Support v3.4
- Ethernet control
CAUSE
Empower with ICF up to v3.4 cannot read the 6890 Plus prerun status during "Waiting for injection".
FIX or WORKAROUND
Upgrade to ICF Support 3.5.
ADDITIONAL INFORMATION
Workaround for ICF 3.4 and lower:
- Use RS 232 for controlling 6890 Plus systems.
- Set up the 6890 Plus correctly in the serial ports tab of node properties.
- Change the baud rate for the 6890 to 19200, and make the "end of command" CR (carriage return) instead of LF.
id99522, 6890A/+GC, 6890N GC