Skip to main content
Waters

Linearity test failure - WKB97258

Article number: 97258

SYMPTOMS

  • Custom linearity test failure after changing from 10-µL to 20-µL loop

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Fixed Loop
  • ACQUITY UPLC Tunable UV Detector

CAUSE

Doubled injection volume causes the highest concentration sample peak to go above the TUV specified range.

FIX or WORKAROUND

Dilute samples.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.