Linearity test failure - WKB97258

Article number: 97258

SYMPTOMS

  • Custom linearity test failure after changing from 10-µL to 20-µL loop

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Fixed Loop
  • ACQUITY UPLC Tunable UV Detector

CAUSE

Doubled injection volume causes the highest concentration sample peak to go above the TUV specified range.

FIX or WORKAROUND

Dilute samples.

ADDITIONAL INFORMATION

 