Linearity test failure - WKB97258
Article number: 97258
SYMPTOMS
- Custom linearity test failure after changing from 10-µL to 20-µL loop
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Fixed Loop
- ACQUITY UPLC Tunable UV Detector
CAUSE
Doubled injection volume causes the highest concentration sample peak to go above the TUV specified range.
FIX or WORKAROUND
Dilute samples.
ADDITIONAL INFORMATION