In a DP2019R1 environment, some items do not display properly when editing the instrument method of FTN - WKB97203
Article number: 97203
SYMPTOMS
- In a DP2019R1 environment, some items do not display properly when editing the FTN instrument method
- When viewing an existing instrument method that contains an H-Class or I-Class FTN module, the following parameter values on the General tab display default values instead of saved values
- Alarm Bands
- Loop Offline
- Load Ahead
- Active Preheater
- Comments
- When retrieving the data, the software uses the values stored in the saved methods
ENVIRONMENT
- DP2019R1 ACQUITY UPLC FTN
- ACQUITY UPLC PLUS FTN
- ACQUITY UPLC Advanced Polymer Chromatography FTN
- Empower
- MassLynx
- Third-party CDS
CAUSE
This is a known issue reported as CRI-908. It was fixed in Waters Driver Pack 2019 Release 2 SR1.
FIX or WORKAROUND
- Make sure the parameter values listed above are correct when viewing and closing the General tab for FTN instrument methods.
- To save the correct parameter values, click Yes in the Save dialog box that appears.
ADDITIONAL INFORMATION
Note: If you open and close the General tab, the Save dialog box appears, even if you want to view it without changing the instrument method.
Recommendations:
- When you open an instrument method to see the parameter value, do not save the method, even if the Save dialog box appears.
- If you change one of the parameter values listed above, make sure the value is correct before saving the method.
- To get a record of parameter values, print a report on the instrument method.
See the following for further information on CRI-908:
https://www.waters.com/waters/suppor...1575&type=RLSN
https://www.waters.com/waters/suppor...0318&type=RLSN
id97203, APH, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, MLYNX, MLYNXV41, SUP, SUPMM, UPARHSMFTN