“Leak Detected” error does not clear with ACQUITY UPLC QSM or BSM - WKB93160
Article number: 93160
SYMPTOMS
- Leak Detected error
- Cleaning the leak sensor does not clear the error
ENVIRONMENT
- ACQUITY UPLC QSM
- ACQUITY UPLC BSM
- ACQUITY APC Isocratic Solvent Manager
CAUSE
- The leak is detected repeatedly due to the solvent buildup in the drip tray.
- The waste tube is not properly installed, the solvent immediately accumulates in the drip tray by the solvent operations such as priming.
- Air lock in waste line.
- Solvent waste line sitting in the liquid within the waste bottle creating back pressure.
- Customer uses restricting barbed fittings or specialist filter caps.
FIX or WORKAROUND
- Remove the leak sensor from the drip tray.
- Wipe off the solvent adhering to the sensor.
- Wipe off the solvent that has accumulated inside the sensor port of the drip tray.
- Check the installation of the waste tube. If it comes straight out of the instrument, plumb it a little downwards so that the solvent will flow under gravity.
- Empty waste bottle.
- Investigate Specialist Waste bottle caps in use.
ADDITIONAL INFORMATION