Skip to main content
Waters

First injection with ACQUITY FTN is inaccurate - WKB89100

Article number: 89100

SYMPTOMS

  • First injection area count is lower than subsequent injections

ENVIRONMENT

  • ACQUITY Arc
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

The septa on vials are not pre-slit.

FIX or WORKAROUND

  1. Determine if a clog is present in the needle (How to test your FTN needle for a partial clog).
  2. Consider use of pre-slit septa.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.