First injection with ACQUITY FTN is inaccurate - WKB89100
Article number: 89100
SYMPTOMS
- First injection area count is lower than subsequent injections
ENVIRONMENT
- ACQUITY Arc
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The septa on vials are not pre-slit.
FIX or WORKAROUND
- Determine if a clog is present in the needle (How to test your FTN needle for a partial clog).
- Consider use of pre-slit septa.
ADDITIONAL INFORMATION