2767 Sample Manager probe hovers above injection ports - WKB89108
Article number: 89108
SYMPTOMS
- Probe hovers above injection ports and wash stations, instead of lowering into injection ports
ENVIRONMENT
- 2767 Sample Manager
- MassLynx
CAUSE
The "Reference Calibrated" check box is empty after calibration of axis.
FIX or WORKAROUND
- Repeat calibrations on autosampler.
- Select the "Reference Calibrated" check box.
- Repeat port washes.
ADDITIONAL INFORMATION
From the Waters 2767 sample manager, injector, and collector installation and maintenance guide (document 71500276704 Rev C):