Skip to main content
Waters

2767 Sample Manager probe hovers above injection ports - WKB89108

Article number: 89108

SYMPTOMS

  • Probe hovers above injection ports and wash stations, instead of lowering into injection ports

ENVIRONMENT

  • 2767 Sample Manager
  • MassLynx

CAUSE

The "Reference Calibrated" check box is empty after calibration of axis.

FIX or WORKAROUND

  1. Repeat calibrations on autosampler.
  2. Select the "Reference Calibrated" check box.
  3. Repeat port washes.

ADDITIONAL INFORMATION

From the Waters 2767 sample manager, injector, and collector installation and maintenance guide (document 71500276704 Rev C):

clipboard_e37dae402c85cdd5c875d3062b3769bf2.png

Not able to find a solution? Click here to request help.