Zp-axis move h/w fault on SM-FTN - WKB85983
Article number: 85983
SYMPTOMS
Zp-axis move h/w fault on SM-FTN
ENVIRONMENT
- ACQUITY SM-FTN I-Class
- ACQUITY SM-FTN H-Class
- ACQUITY Arc Sample Manager FTN-R
CAUSE
Missing vial in V:1 position on tray in SM-FTN.
FIX or WORKAROUND
Put a vial into V:1 position in SM-FTN or inject from plate.
ADDITIONAL INFORMATION