ACQUITY UPLC SM-FTN - Failing system leak test (including needle and seal) - WKB85864

Article number: 85864

SYMPTOMS

  • Noisy and sloping baseline
  • Dynamic leak test (to column end point, including the needle and seal) fails
  • Dynamic leak test (to column end point, excluding the needle and seal) passes

ENVIRONMENT

  • ACQUITY UPLC H-class Sample Manager - FTN
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC I-Class Sample Manager - FTN
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY Arc Sample Manager FTN-R

CAUSE

Excessive leak on the extension loop.

FIX or WORKAROUND

  1. Replace extension loop (What are the part numbers for H-Class Sample Manager Flow Through Needle extension loops? - WKB6746)
  2. Verify dynamic leak test (to column end point, including the needle and seal) now passes

ADDITIONAL INFORMATION