Noisy Baseline and Bad Chromatography on 2998 - WKB85306

Article number: 85306

SYMPTOMS

  • occasionally noisy baseline 
  • poor chromatography

ENVIRONMENT

  • 2998 Photodiode Array Detector (Platinum)
  • ACQUITY UPLC PDA Detector

CAUSE

Air conditioning affects internal temperature

FIX or WORKAROUND

  1. Move the system to another place where the A/C doesn't blow directly onto the detector