ACQUITY UPLC Sample Manager - FTN: "Needle/Seal move limit exceeded"- WKB85308
Article number: 85308
SYMPTOMS
- Error "Fatal: Needle/Seal move limit exceeded"
- Error can be temporarily cleared
- Characterization of the needle seal giving seal constant out of range. Expected value is 8-11 N/mm.
ENVIRONMENT
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC H-class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The inject station is not assembled correctly: prongs on the bottom of the locking nut (seal port) do not align with the spring (see Figure 1).
FIX or WORKAROUND
- Dismantle the inject station and rebuild it with the spring correctly aligned
- Re-characterize the needle seal and perform the needle seal readiness test
ADDITIONAL INFORMATION