Which two wires on the relay cable between the ACQUITY Sample Manager and a Sciex mass spectrometer (MS) need to be connected to Inject Start Out to tigger the MS to start? - WKB83642
Article number: 83642
ENVIRONMENT
- ACQUITY UPLC
- Sciex mass spectrometer
ANSWER
The pair of wires on the relay cable required to trigger the Sciex MS are the white one with black stripes and the green one with black stripes. They are connected to Inject Start Out on the back of the ACQUITY Sample Manager. In addition, the red wire with black stripes must be shorted against the orange wire with black stripes.