2695 Separations Module and 2489 UV/Vis Detector do not communicate with Empower - WKB83571
Article number: 83571
SYMPTOMS
- Instrument does not populate in Waters DHCP Server Configuration
- 2695 Separations Module is IEEE
- 2489 UV/Vis Detector is Ethernet
ENVIRONMENT
- Empower 3
- Empower 2
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- 2695 Separations Module
- 2489 UV/Vis Detector
CAUSE
Network adaptors are not bridged.
FIX or WORKAROUND
- Bridge the two network adaptors in Windows.
- Power-up the Ethernet instrument. (In DHCP, the IP address, MAC address, and Type show up, but the name does not.)
- From the front panel of the instrument, specify the serial numbers, and then reboot.
- The serial number now populates in DHCP.
- LAC/E node property shows Yes in the Status OK? column.