Alliance HPLC system does not hold pressure - WKB83547

Article number: 83547

SYMPTOMS

  • Alliance not holding pressure
  • High pressure ripple observed, 70 psi - 200 psi, delta at 3,500 psi

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Inlet tubing possibly clogged.

FIX or WORKAROUND

  1. Power-off the Alliance or turn off the GPV using the diagnostics.
  2. Replace the Standard Flow Inlet Tubing Assembly (WAT270986).
  3. Turn the Alliance back ON or press "OK" to turn on the GPV.

ADDITIONAL INFORMATION

 

id83547, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, gradient proportioning valve, pressure fluctuation

Not able to find a solution? Click here to request help.