How to access the dry and wet prime functions in the Alliance 2695 - WKB83536
Article number: 83536
OBJECTIVE or GOAL
Access the dry and wet prime functions in the Alliance 2695.
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- Empower
PROCEDURE
- Exit out of the Empower Run Samples page.
- Press Menu/Status on the 2695 to display the Status screen.
- Press the Direct Function screen key. A list of direct functions appears; look for dry prime and wet prime.