How to install an extension loop on ACQUITY H-Class FTN? - WKB8352
Article number: 8352
OBJECTIVE or GOAL
How to install an extension loop on ACQUITY H-Class FTN.
ENVIRONMENT
- ACQUITY H-Class FTN (flow-through needle), Sample Manager
- ACQUITY UPLC I-Class PLUS
PROCEDURE
To install or replace an extension loop:
- Power-off the sample manager.
- Open the fluidics compartment door.
- Using the 1/4-inch open-end wrench, unscrew the fitting of the needle tubing or extension loop from port of the injection valve.

Needle tubing (or extension loop tubing, if installed)
- Fully insert the extension-loop tubing into port 4 of the injection valve, and then thread the fitting into the port.
- Holding the extension loop tubing against the bottom of the port, finger-tighten the extension-loop fitting.
- Using the 1/4-inch open-end wrench, tighten the extension-loop fitting an additional 1/6-turn (for existing fittings) or 3/4-turn (for a new fitting).
Note: Failure to properly bottom the extension loop tubing in the port can result in carryover or poor chromatography.
- Screw the sample needle's fitting into the extension-loop union, and then use the 1/4-inch open-end wrench to tighten the fitting 3/4-turn beyond finger-tight (for a new fitting) or 1/6-turn beyond finger-tight (for existing fittings).
Note: Failure to properly bottom the sample needle tubing in the extension-loop union could result in carryover or poor chromatography.

Extension loop Union
Extension loop tubing
Extension loop fitting
- Close the fluidics compartment door.
- Power-on the sample manager.
- Modify the volume of the extension loop according to the following instructions.
To modify the needle or extension-loop volume setting:
- In the console, select the sample manager, and then click Configure > Volumes.
- In the Volume Configuration dialog box, select the appropriately sized needle or extension loop size from the list, and then click OK.
Note: If the sample manager is not fitted with an extension loop, select 0 as the loop size
ADDITIONAL INFORMATION
For part numbers on extension loops, see WKB6746
Flow Through Needle Overview and Maintenance guide