High system pressure on ACQUITY UPLC Sample Manager - Flow Through Needle - WKB81819
hSYMPTOMS
- High pressure
- System overpressure error
- Issue observed when in inject position, not in load position
ENVIRONMENT
- ACQUITY Arc
- ACQUITY Arc Bio
- ACQUITY Arc FTN-R
- ACQUITY Arc Bio FTN-R
- ACQUITY Arc Premier FTN-R
- ACQUITY UPLC Sample Manager - FTN
- ACQUITY UPLC I-Class FTN
CAUSE
Blockage, clog, or bend of needle and/or needle seat
FIX or WORKAROUND
- Replace the needle and needle seat.
- Calibrate needle z axis
Part numbers for replacement needles (these parts include the needle seat):
700005279 - 30-µL stainless steel needle
700009578 - 30-µL MP35N needle (Bio)
700005215 - 15-µL stainless steel needle
700005421 - 15-µL MP35N needle (Bio)
ADDITIONAL INFORMATION
These issues are often caused by use of applications that employ relatively high concentrations of non-volatile buffer salts in the mobile phases—typically applications taken from pharmacopoeias (USP/Ph.Eur./JP, others).
When using such procedures, it is recommended that the system be flushed and cleaned regularly to remove excess buildup of buffer salts, which can result in pressure issues at various locations within the system.