Skip to main content
Waters

Accumulator pump is unable to compress during leak test - WKB80402

Article number: 80402

SYMPTOMS

  • Accumulator is unable to compress during leak test.
  • No solvent comes out of waste line during priming.
  • No solvent comes out of primary transducer to accumulator tubing.

ENVIRONMENT

  • ACQUITY UPLC Classic
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY Arc

CAUSE

The intelligent intake valve (I2V) cartridge is faulty.

FIX or WORKAROUND

Replace the intelligent intake valve (I2V) cartridge (Part Number 700003563).

If the problem persists, replace the intelligent intake valve (Part Number 700003557).

ADDITIONAL INFORMATION

 

id80402, A-30SDM, UPAQSM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM, UPPDAARCB, UPPQSM, UPQSM, UPQSM+, UPQSMBIO

Not able to find a solution? Click here to request help.