"Preheater configured but not detected" error for ACQUITY UPLC H-Class using a CH-30A with a CM-A - WKB80265
Article number: 80265
SYMPTOMS
- "Preheater configured but not detected" error when configuring a CH-30A in a system with a CM-A
ENVIRONMENT
- ACQUITY UPLC H-Class
- Empower 3 FR4
- ICOP 1.69 for the CM and 1.70 for the SM
CAUSE
Method is incorrectly configured in Empower to tell the software to ignore the empty APH slots on the CM-A.
FIX or WORKAROUND
- In the console for the column manager instrument configuration, choose the number of columns required.
- Go to preheater configuration on the column manager console page and choose Active, Disabled for the APH positions not used on the column manager.
- Go to the inlet method for the column manager and, for the Active preheater, choose Disabled or Use console configuration (either works).
- Go to the console sample manager page and, in the preheater configuration, select Active.
- In the inlet method for the sample manager, change the active preheater option to Enabled.
ADDITIONAL INFORMATION
The setup above also works if a CM-Aux is present. However, the number of columns will be different compared with the screenshot in the attachment.