ACQUITY UPLC detector is leaking - WKB80257
Article number: 80257
SYMPTOMS
- Leaking from the inlet port of the flow cell
ENVIRONMENT
- ACQUITY UPLC PDA Detector
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC FLR Detector
CAUSE
Damaged finger-tight PEEK fitting
FIX or WORKAROUND
Replace the PEEK fitting.
ADDITIONAL INFORMATION
For part numbers of UPLC PEEK fittings/assemblies, see article What is the part number for the inlet tubing to an ACQUITY PDA detector? - WKB10850