Agilent A1100 LC causes instrument failure - WKB7978
Article number: 7978
SYMPTOMS
- On the Instruments tab, in node properties, the A1100 is listed as NO in the OK? column
- Instrument Failure error when setting up an instrument method
- IP address manually added to the DHCP
ENVIRONMENT
- Agilent 1100:
- G1323B Handheld Control Module
- G1312A Binary Pump
- G1329A Standard Autosampler
- G1316A Thermostatted Column Compartment
- G1365B Multi-Wavelength Detector (MWD)
- G1846A (JetDirect 400N J4100A) network card
- Empower 3
- Agilent LC Instrument Control Software version 1.06
CAUSE
Bad Agilent G1846A (JetDirect 400N J4100A) network card
FIX or WORKAROUND
Replace the Agilent G1846A (JetDirect 400N J4100A) network card.