Skip to main content
Waters

Corona CAD detector control for Empower will not go to next injection after the first - WKB79813

Article number: 79813

SYMPTOMS

  • Sample set has 15 injections with a run time of 30 minutes.
  • The first injection runs fine, but fails to go to the next injection.

ENVIRONMENT

  • Empower 3
  • Empower 2
  • Empower 3 Personal
  • Empower 3 Workgroup
  • Empower 3 Enterprise
  • Thermo Corona CAD Detector

CAUSE

Run time not set in the instrument method.

FIX or WORKAROUND

  1. Open the instrument method.
  2. Select the CAD detector.
  3. Specify the Stop Time to match the Run Time in the Sample Set Method.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.