Corona CAD detector control for Empower will not go to next injection after the first - WKB79813
Article number: 79813
SYMPTOMS
- Sample set has 15 injections with a run time of 30 minutes.
- The first injection runs fine, but fails to go to the next injection.
ENVIRONMENT
- Empower 3
- Empower 2
- Empower 3 Personal
- Empower 3 Workgroup
- Empower 3 Enterprise
- Thermo Corona CAD Detector
CAUSE
Run time not set in the instrument method.
FIX or WORKAROUND
- Open the instrument method.
- Select the CAD detector.
- Specify the Stop Time to match the Run Time in the Sample Set Method.