Communication failure with e-SAT/IN module - WKB7971
Article number: 7971
SYMPTOMS
- Instrument failure when trying to set up an instrument method
- In Node Properties, the e-SAT/IN is listed as NO in the OK? column on the Instruments tab
- Rebooting all components, including the PC, does not correct the communication issue
- Alternative IP addresses do not correct the communication issue
- Cannot ping to the systems, but can ping to the NIC card
ENVIRONMENT
- Empower 3 Personal
CAUSE
Windows Firewall is turned on without any exclusions.
FIX or WORKAROUND
- Make exclusions as listed in the Empower 3 (Feature Release 3) Installation, Configuration, and Upgrade Guide.
Or
- Turn the Windows Firewall off.