Peak broadening and tailing - WKB79673
Article number: 79673
SYMPTOMS
- Broad peaks
- Large tailing factor
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- ACQUITY UPLC
CAUSE
- Column inlet and/or outlet tubing diameter incorrect, 0.020" tubing connecting column inlet and/or outlet.
- Dead volume at column inlet because of improper capillary intrusion length.
- Leakage at the inlet of the column.

FIX or WORKAROUND
- Replace column inlet and/or outlet tubing.0.009" tubing connects to column inlet and outlet.
- Adjust the capillary intrusion into the column to a proper length.
- Tighten column inlet capillary fittings.

ADDITIONAL INFORMATION
Alliance (old) - 1/16 in. (1.6mm) OD x 0.009 in. (0.23 mm) ID x 10 ft (3.0 m) length (WAT026973)
Alliance (new) - ASSY, TUBE, PEEK, .062 X .005 X 20 L (430000922)
HPLC - Compression Screws and Ferrules, 5/Pkg (WAT025604)
UPLC Classic - Column Stabilizer Kit, 150 mm (205000494)
H-Class UPLC - APH, SS, 12.5 LG (205001774)
I-Class UPLC - APH, SS, 12.5 LG, .003 ID (205000894)
