Skip to main content
Waters

Error in ACQUITY UPLC H-Class FTN: Sample temperature range error 0.0 - WKB7960

Article number: 7960

SYMPTOMS

  • In Empower Run Samples, immediately after a run is started, error "Sample temperature range error 0.0" appears

ENVIRONMENT

  • ACQUITY UPLC H-Class FTN
  • Empower 3

CAUSE

The method does not have the ACQUITY UPLC H-Class FTN listed.

FIX or WORKAROUND

  1. In Empower, create a new instrument method including the ACQUITY UPLC H-Class FTN. Specify a temperature in the General tab of the FTN (see screen shot below).
  2. Save the method.
  3. Run samples.

FTN general tab.PNG

ADDITIONAL INFORMATION

 

id7960, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.