Chromatographic data shows both high backgound signal and discrete contamination peaks in the trace for data acquired on an ACQUITY UPLC H-Class and Xevo TQD - WKB79501
Article number: 79501
SYMPTOMS
- High background signal
- Discreet peaks in the trace that should not be there
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
- ACQUITY UPLC
- SQD2
- Xevo TQD
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-S cronos
CAUSE
System is contaminated both in the solvent lines and in the injector. Source of contamination unknown
FIX or WORKAROUND
- Replace chromatographic column with a union or set the the column position for the Column Manager to bypass
- On the MS Tune page, set the flow state for the fluidics valve to waste
- Prepare a solution of Water/Methanol/Acetonitrile/Isopropanol, 25/25/25/25 (v/v/v/v). Prepare at least 1 liter.
- Fill 10 2 mL sample vials with this solution. Place vials in sample tray. Put tray into Sample Manager
- Put all solvent, wash, and purge (if applicable) lines for the Solvent Manager and Sample Manager into the bottle containing the solution
- Prime all solvent, wash and purge lines with this solution.
- Create an inlet method for an isocratic run with a run time of 3 minutes
- For QSM, method should be 25% for each solvent line
- For BSM method should be 50% for each solvent line
- Set up a batch in the Sample List (MassLynx) or Sample Set (Empower) that will make 5 injections from each vial (50 injections total) using the largest allowable injection volume for the current configuration.
- Flush the system, all lines in the Sample Manager and Solvent Manager with 100% methanol.
- Flush and prime the lines with the appropriate solvents for the normal assay run.
- Run Blank injections to see if contamination has been cleared or if it still persists.
- If contamination persists, repeat procedure.
- Repeat procedure up to 3 times. If issue persists, contact the Expert Center in your geography.
ADDITIONAL INFORMATION
id79501, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO