What is the maximum injection volume on a ACQUITY Flow Through Needle (FTN)? - WKB7947

Article number: 7947

ENVIRONMENT

  • ACQUITY UPLC H-Class FTN
  • ACQUITY UPLC I-Class FTN

ANSWER

The maximum allowed injection is as follows: 

  • If no extension loop is installed, the extension loop size is 0, so the maximum volume is the needle size minus 5 uL.
  • If an extension loop is installed, then the maximum volume is 100% of the installed extension loop.

ADDITIONAL INFORMATION