ACQUITY UPLC Sample Manager error: Failed to Calibrate system volume when calibrating the VDD - WKB79300
Article number: 79300
SYMPTOMS
- Error "Failed to Calibrate system volume" when characterizing the needle and loop
- Error "Failed to Calibrate system volume" when calibrating the VDD
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Defective injection valve cartridge
FIX or WORKAROUND
- Install a new injection valve cartridge (700002765).
- Recalibrate VDD.
- Characterize needle and loop volumes.
