Sloping baseline with 2414 RID after mobile phase change - WKB79233
Article number: 79233
SYMPTOMS
- Noisy baseline
ENVIRONMENT
- 2414 Refractive Index Detector
CAUSE
Reference cell was not purged.
FIX or WORKAROUND
Purge the reference cell with fresh eluent.