Area counts differ between an Alliance 2695 HPLC with 2996 detector and the same with 2998 detector - WKB7884

Article number: 7884

SYMPTOMS

  • Area counts are different on an Alliance 2695 HPLC with 2996 detector and an Alliance 2695 HPLC with 2998 detector when using the same column, mobile phase, and sample
  • Peak shape is comparable
  • RT is the same
  • Values are reproducible

ENVIRONMENT

FIX or WORKAROUND

  1. Install the Analytical Flow Cell Kit [205000399].
  2. Rerun the sample.

ADDITIONAL INFORMATION

 

id7884, 2996, 2998, 2998B, 2998SFC, A-10PDA, eluent