Area counts differ between an Alliance 2695 HPLC with 2996 detector and the same with 2998 detector - WKB7884

Article number: 7884

SYMPTOMS

  • Area counts are different on an Alliance 2695 HPLC with 2996 detector and an Alliance 2695 HPLC with 2998 detector when using the same column, mobile phase, and sample
  • Peak shape is comparable
  • RT is the same
  • Values are reproducible

ENVIRONMENT

FIX or WORKAROUND

  1. Install the Analytical Flow Cell Kit [205000399].
  2. Rerun the sample.

ADDITIONAL INFORMATION

 

id7884, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, 2996, 2998, 2998B, 2998SFC, A-10PDA, A-10SSM, eluent