Skip to main content
Waters

Random pressure drops with Alliance 2695 - WKB78171

Article number: 78171

SYMPTOMS

  • Random pressure drops
  • Leak test passes

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Broken fitting on the tubing that runs from the manifold to the transducer

FIX or WORKAROUND

  1. Replace the tubing and the manifold.
  2. Run the static leak test.

ADDITIONAL INFORMATION

The primary-to-transducer part will come up as "Accumulator to Bottom transducer", but this is the correct part number.

Not able to find a solution? Click here to request help.