Noisy baseline on 2489 UV/Vis detector - WKB78148
Article number: 78148
SYMPTOMS
- Noisy baseline
- Lamp is new
ENVIRONMENT
- 2489 UV/Vis Detector
CAUSE
Incorrect inlet/waste configuration on the 2489 detector
FIX or WORKAROUND
Connect the the flow cell inlet to the column outlet and the flow cell outlet to waste.
ADDITIONAL INFORMATION
Due to improper inlet/waste tubing configuration, the flow cell might be damaged. Replace if necessary.
id78148, 2489, 2489B, 2489SFC, A-10UV