SQT test injections fail on UPLC H-Class - WKB78143
Article number: 78143
SYMPTOMS
- Failure to meet 1.8 AU peak height requirement on SQT test injections
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC PDA Detector
- SystemsQT (SQT)
CAUSE
- The column outlet - flow cell inlet tubing is not the original (0.005 ID)
FIX or WORKAROUND
Replace the column outlet - flow cell inlet tubing (0.004 ID, PDA inlet).