Retention times shift out in ACQUITY UPLC H-Class - WKB76868

Article number: 76868

SYMPTOMS

  • Pressure is stable
  • Pressure is lower than normal
  • Retention time shifts during sample set
  • The dynamic leak end point vent valve test passes
  • The dynamic leak test with column end point passes

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Arc
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

Leaking needle seal (405011492)

FIX or WORKAROUND

  1. Replace the needle seal.
  2. Purge the fluidics.
  3. Perform needle seal characterization.
  4. Perform a dynamic leak test on the injector pod.

ADDITIONAL INFORMATION

Consider replacing the needle as well. A blunt tip can cause premature seal failure.

The part number for the standard 15-µL needle with vespel seat is 700005215.

Video for replacement of needle and vespel seat (needle seal)

Needle replacement written procedure

id76868, A-30SM, DLT, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARHSMFTN, UPARQSM, UPARSMFTN, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPLCCOLHTR, UPNBINARY, UPPDAARCB, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR