Article number: 76862

SYMPTOMS

  • High pressure readback on instrument console
  • Higher than expected pressure when column disconnected
  • Pressure drops when tubing between the Sample Manager injector outlet and the Column Manager inlet is removed

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class

CAUSE

Blockage in the tubing between the Sample Manager injector outlet and the Column Manager inlet

FIX or WORKAROUND

Replace capillary from Sample Manager to Column Manager

For I-Class use part number 700008869 - Welded Tube Assy, .003 ID x 14.5 long

For H-Class use part number 700005476 - Welded Tube Assy, .004 ID x 14.5 long

ADDITIONAL INFORMATION