Arc column position leak test failure - WKB76774

Article number: 76774

SYMPTOMS

  • Leak test (vent valve) passes
  • Leak test (column position) fails

ENVIRONMENT

  • ACQUITY Arc

CAUSE

Fitting deterioration

FIX or WORKAROUND

Check and replace the fitting at the leaking point.

Part numbers:

  700010687 - Vent Valve P1

  700010686 - Vent Valve P2

  700010685 - Vent Valve P4

  700010685 - Vent Valve P5

  700010703 - Vent Valve FTN-R

ADDITIONAL INFORMATION