Batch immediately goes to "Queue is paused" when sample run is started in MassLynx for an Autopurification system - WKB75154
Article number: 75154
SYMPTOMS
- Queue status immediately goes to "Queue is paused" when run is started
- Reset System does not resolve
- Found: MS in Standby
- Found: LC flow not on
- Found: Fraction collector, Collection control, System Error
- Found: Cannot activate the fraction collector
- Found: Reset Comm from Collection control and Inlet method does not resolve
- Found: Reset Injector from Inlet method does not resolve
- Found: No bed layout displayed in Collection Control
ENVIRONMENT
- MassLynx
- Autopurification System
- 2767 Sample Manager
- 3767 Sample Manager
CAUSE
Communication issue between 2767 or 3767 and MassLynx PC
FIX or WORKAROUND
- Close MassLynx and Collection Control.
- Reboot 2767 hardware.
- Open MassLynx and Collection Control.
- Activate fraction collector.
- Reset beds from Collection Control.
ADDITIONAL INFORMATION